Performance Specifications |
Dimension | 360mm*310mm*120mm |
Weight | 5.3kg(excluding accessories) |
Power Supply | Voltage AC100~240V, 50/60HZ |
Display | 12.1" color TFT LED (0-90° rotation) Resolution 1024*768 pixels |
Battery (Built-in) | Type Rechargeable lithium battery 14.8V/2200mAh |
Charge Cycle | ≥500 times |
Working time | 2 hours |
Recorder (Built-in) |
Method | Thermal printer |
Paper | Z-fold, thermosensitive |
Paper width | 112 mm (4.41 in)/150 mm (5.91 in) |
Printing speed | 1/2/3cm/min |
Trace | Max. 4 tracks |
-Recording way | Real-time Recording, Review Printing |
Alarm |
|
Level | Low, medium and high |
Indication | Auditory, Visual, Delay of FHR Alarm |
Input Device |
|
-Touch screen (option) |
|
-Knob (Standard) |
|
-Keypad input (Standard) |
|
Mouse/Keyboard input (option) |
|
-USB port for patient data/NST report backup (option) |
|
System Output & Extensible Interface |
|
-Ethernet Network | 1 standard RJ45 socket |
-USB Port | 2 |
-Video Output | 1 HDMI port (option) |
Trend & Reviewing |
|
-Cardiotocography (CTG) |
|
Waveform Review | 8640 hours |
-Vital Signs |
|
Trend | 2160 hours |
Waveform Review | 2 hours |
NIBP Review | 2000 groups |
-Alarm Review | 2000 groups |
Safety |
|
-IEC60601-1 Approved, CE marking according to MDD93/42/EEC |
|
-With reference to RoHS Directive 2011/65/EU recasting |
|
-Ultrasound (FHR) |
|
-Transducer type | 12-Crystal ultrasound transducer |
-Measure method | Wide beam pulse doppler |
-FHR Measurement range |
|
30 to 240 bpm (US Standard) |
|
50 to 210 bpm (Europe Standard) |
|
Accuracy | ±1 bpm |
Intensity | <3 mW/cm2 |
-Ultrasound frequency | 1MHz±3% |
Resolution | 1 bpm |
TOCO |
|
-Measurement method | Strain gauge sensor element |
-Measure range | 0 ~ 100 units |
Resolution | 1 unit |
-Nonlinear deviation | ≤±8% |
-Baseline setting | 0,5,10,15, 20 can be selected |
-Auto offset correction | 3 second after connecting the transducer |
-Auto zero adjust | TOCO value is set to zero negative following a negative Measure value for 5 seconds |
Fetal Movement Recording |
|
-Automatic fetal movement (AFM) | pulsed doppler ultrasound |
-Automatic fetal movement autograph |
|
-Manual fetal movement marker (MFM) |
|
Maternal ECG |
|
-Lead mode | 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V |
A63Gain | 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, Auto |
Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s, 50 mm/s |
Accuracy | ±1bpm/ ±1%, whichever is greater |
Resolution | 1 bpm |
Leakage Current < 10 μA |
|
CMRR | Diagnostic Mode >89 dB |
| Monitor Mode >105 dB |
| Surgery Mode >105 dB |
Bandwidth | Surgery 1 ~ 25 Hz |
| Monitor 0.5 ~ 40 Hz |
| Diagnostic 0.05 ~ 120 Hz |
Maternal Heart Rate |
|
-Measure range | Adult:0bpm; 10 ~ 300 bpm |
| Neonate:0bpm; 10 ~ 350 bpm |
Resolution | 1 bpm |
Accuracy | ± 1% or ± 1bpm, whichever is greater |
ST Measurement |
|
Range | -2.0 ~ +2.0 mV |
Accuracy | -0.8mV~+0.8mV:±0.02mV or ±10%, whichever is greater |
-Other range | unspecified |
Resolution | 0.01mV |
Maternal Respiration |
|
Method | Impedance between RA-LL, RA-LA |
Gain | ×0.25, ×0.50, ×1, ×2, ×4 |
-Respiration Rate | 0bpm, 6 ~ 150 bpm |
Resolution | 1 bpm |
Accuracy | ±2 bpm or ±2%, whichever is greater |
-Apnea Alarm | 10 ~ 60 s |
Maternal SpO2 (Digital Technic) |
|
-Measurement Range | 0 ~ 100 % |
Resolution | 1% |
Accuracy | ±2% (70% ~ 100%) |
±3 % (35% ~ 69%) |
|
Unspecified (0 ~ 34%) |
|
Pulse Rate |
|
Range | 25~250bpm |
Accuracy | ±1% or ±1 bpm, whichever is greater |
Resolution | 1bpm |
Maternal NIBP (MNIBP) |
|
Method | Oscillometric |
-Measure mode Manual, Auto, STAT |
|
-Measure Interval in AUTO Mode 1~480 min |
|
-STAT mode cycle time |
|
-Keep 5 minutes, at 5 seconds interval |
|
-Measure and Alarm Range |
|
Systolic | 30~270 mmHg |
Diastolic | 10~220 mmHg |
Mean | 20~235 mmHg |
-Static pressure accuracy ±3mmHg |
|
Resolution | 1mmHg |
Accuracy | Maximum Mean error ±5mmHg |
Maximum Standard deviation 8mmHg |
|
-Over pressure Protection | Dual protection via software & hardware |
Maternal Temperature |
|
Technique | Thermistor probe (2.25K) |
-Measuring and Alarm Range | 0.0 ℃~ 50 ℃ (32℉~ 122℉) |
Unit | Celsius (℃), Fahrenheit ( ℉) |
Resolution | 0.1℃ |
Accuracy | ±0.1℃ |
NST Report/FHR Interpretation (Option) |
|
-Based on Fischer Evaluation Method (monitoring time≥20 Min) |
|
-Average Baseline FHR /Number of Acceleration /Number of Deceleration /Number of Small Acceleration /Amplitude Variation /Number |
|
of PV |
-Automatic Score Result | Score 8-10 (Excellent) |
Score5-7 (Suspicious) |
Score ≤4 ( Fetus Hypoxia) |