| Performance Specifications |
| Dimension | 360mm*310mm*120mm |
| Weight | 5.3kg(excluding accessories) |
| Power Supply | Voltage AC100~240V, 50/60HZ |
| Display | 12.1" color TFT LED (0-90° rotation) Resolution 1024*768 pixels |
| Battery (Built-in) | Type Rechargeable lithium battery 14.8V/2200mAh |
| Charge Cycle | ≥500 times |
| Working time | 2 hours |
| Recorder (Built-in) |
| Method | Thermal printer |
| Paper | Z-fold, thermosensitive |
| Paper width | 112 mm (4.41 in)/150 mm (5.91 in) |
| Printing speed | 1/2/3cm/min |
| Trace | Max. 4 tracks |
| -Recording way | Real-time Recording, Review Printing |
| Alarm |
|
| Level | Low, medium and high |
| Indication | Auditory, Visual, Delay of FHR Alarm |
| Input Device |
|
| -Touch screen (option) |
|
| -Knob (Standard) |
|
| -Keypad input (Standard) |
|
| Mouse/Keyboard input (option) |
|
| -USB port for patient data/NST report backup (option) |
|
| System Output & Extensible Interface |
|
| -Ethernet Network | 1 standard RJ45 socket |
| -USB Port | 2 |
| -Video Output | 1 HDMI port (option) |
| Trend & Reviewing |
|
| -Cardiotocography (CTG) |
|
| Waveform Review | 8640 hours |
| -Vital Signs |
|
| Trend | 2160 hours |
| Waveform Review | 2 hours |
| NIBP Review | 2000 groups |
| -Alarm Review | 2000 groups |
| Safety |
|
| -IEC60601-1 Approved, CE marking according to MDD93/42/EEC |
|
| -With reference to RoHS Directive 2011/65/EU recasting |
|
| -Ultrasound (FHR) |
|
| -Transducer type | 12-Crystal ultrasound transducer |
| -Measure method | Wide beam pulse doppler |
| -FHR Measurement range |
|
| 30 to 240 bpm (US Standard) |
|
| 50 to 210 bpm (Europe Standard) |
|
| Accuracy | ±1 bpm |
| Intensity | <3 mW/cm2 |
| -Ultrasound frequency | 1MHz±3% |
| Resolution | 1 bpm |
| TOCO |
|
| -Measurement method | Strain gauge sensor element |
| -Measure range | 0 ~ 100 units |
| Resolution | 1 unit |
| -Nonlinear deviation | ≤±8% |
| -Baseline setting | 0,5,10,15, 20 can be selected |
| -Auto offset correction | 3 second after connecting the transducer |
| -Auto zero adjust | TOCO value is set to zero negative following a negative Measure value for 5 seconds |
| Fetal Movement Recording |
|
| -Automatic fetal movement (AFM) | pulsed doppler ultrasound |
| -Automatic fetal movement autograph |
|
| -Manual fetal movement marker (MFM) |
|
| Maternal ECG |
|
| -Lead mode | 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V |
| A63Gain | 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, Auto |
| Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s, 50 mm/s |
| Accuracy | ±1bpm/ ±1%, whichever is greater |
| Resolution | 1 bpm |
| Leakage Current < 10 μA |
|
| CMRR | Diagnostic Mode >89 dB |
| Monitor Mode >105 dB |
| Surgery Mode >105 dB |
| Bandwidth | Surgery 1 ~ 25 Hz |
| Monitor 0.5 ~ 40 Hz |
| Diagnostic 0.05 ~ 120 Hz |
| Maternal Heart Rate |
|
| -Measure range | Adult:0bpm; 10 ~ 300 bpm |
| Neonate:0bpm; 10 ~ 350 bpm |
| Resolution | 1 bpm |
| Accuracy | ± 1% or ± 1bpm, whichever is greater |
| ST Measurement |
|
| Range | -2.0 ~ +2.0 mV |
| Accuracy | -0.8mV~+0.8mV:±0.02mV or ±10%, whichever is greater |
| -Other range | unspecified |
| Resolution | 0.01mV |
| Maternal Respiration |
|
| Method | Impedance between RA-LL, RA-LA |
| Gain | ×0.25, ×0.50, ×1, ×2, ×4 |
| -Respiration Rate | 0bpm, 6 ~ 150 bpm |
| Resolution | 1 bpm |
| Accuracy | ±2 bpm or ±2%, whichever is greater |
| -Apnea Alarm | 10 ~ 60 s |
| Maternal SpO2 (Digital Technic) |
|
| -Measurement Range | 0 ~ 100 % |
| Resolution | 1% |
| Accuracy | ±2% (70% ~ 100%) |
| ±3 % (35% ~ 69%) |
|
| Unspecified (0 ~ 34%) |
|
| Pulse Rate |
|
| Range | 25~250bpm |
| Accuracy | ±1% or ±1 bpm, whichever is greater |
| Resolution | 1bpm |
| Maternal NIBP (MNIBP) |
|
| Method | Oscillometric |
| -Measure mode Manual, Auto, STAT |
|
| -Measure Interval in AUTO Mode 1~480 min |
|
| -STAT mode cycle time |
|
| -Keep 5 minutes, at 5 seconds interval |
|
| -Measure and Alarm Range |
|
| Systolic | 30~270 mmHg |
| Diastolic | 10~220 mmHg |
| Mean | 20~235 mmHg |
| -Static pressure accuracy ±3mmHg |
|
| Resolution | 1mmHg |
| Accuracy | Maximum Mean error ±5mmHg |
| Maximum Standard deviation 8mmHg |
|
| -Over pressure Protection | Dual protection via software & hardware |
| Maternal Temperature |
|
| Technique | Thermistor probe (2.25K) |
| -Measuring and Alarm Range | 0.0 ℃~ 50 ℃ (32℉~ 122℉) |
| Unit | Celsius (℃), Fahrenheit ( ℉) |
| Resolution | 0.1℃ |
| Accuracy | ±0.1℃ |
| NST Report/FHR Interpretation (Option) |
|
| -Based on Fischer Evaluation Method (monitoring time≥20 Min) |
|
| -Average Baseline FHR /Number of Acceleration /Number of Deceleration /Number of Small Acceleration /Amplitude Variation /Number |
|
| of PV |
| -Automatic Score Result | Score 8-10 (Excellent) |
| Score5-7 (Suspicious) |
| Score ≤4 ( Fetus Hypoxia) |